Sentra Biosains Dinamika

Berthold - LB 148 Hand-Foot Monitor with Back of Hand Measurement, Mobil (wheels and handle)

Catalog No. 56542 Category

PERSONAL CONTAMINATION MONITOR

Radionuclides are used in many areas of research, nuclear medicine and natural sciences. Wherever open radioactive materials are handled, contamination measurements by hand, feet and clothing are required in accordance with radiation protection guidelines.

The Personal Contamination Monitors LB 147 and LB 148 are contamination monitors designed for use in the radiation protection field to detect contamination caused by alpha and beta/gamma radiation. Low-maintenance thin-layer ZnS (Ag) scintillators are used as detectors. The operation is via a touch panel on a graphical display.

LB 148

  • Simultaneous back-of-hand measurement with automatically tracked back-of-hand probe for optimum measurement geometry
  • Self-adjusting back of hand detectors
  • Time efficient measurement
  • Dual password hierarchy
  • Simple operation with graphic display and touch panel

Brand

Berthold

 LB 147LB 147 SlimLB 148Alpha Sense
ZnS detector
Number of detectors44645022
Removable hand probe for clothing measurements
Weight (approx. in kg)35354535 / 45
Dimensions (without alarm tower)65 x 125 x 8045 x 125 x 9085 x 125 x 82240 x 140 x 110
(W x H x D) in mm    
InterfacesUSB, Ethernet, RS232, RS485USB, Ethernet, RS232, RS485USB, Ethernet, RS232, RS485USB, Ethernet, RS232, RS485
Mobile set (fixed rollers)opt.opt.opt. / ✓
     
     
Hand Detectors    
     
DetectorZnS(Ag) scintillator with light-tight Hostaphan foil  
Dimensions entrance window150 mm x 230 mm   
Senstive area345 cm2   
Typical background    
α-channelapprox. 0.1 cps   
β-channelapprox. 15 cps   
     
     
Typical efficiencies (ISO 7503-1)    
 LB 147 / LB 148Alpha Sense Version  
14C20 %0,25  
36Cl0,490,7  
90Sr / 90Y0,520,66  
241Am0,330,43  
     
     
Typical detection limits (ISO 7503-1) with 10 s measuring time in Bq/cm2   
 LB 147 / LB 148Alpha Sense Version  
14C0.6490.460  
36Cl0.1350.078  
90Sr / 90Y0.1260.088  
241Am0.0260.019  
     
     
Foot Detectors    
     
DetectorZnS(Ag) scintillator with light-tight Hostaphan foil  
Dimensions entrance window150 mm x 370 mm   
Senstive area555 cm2   
Typical background    
α-channelapprox. 0.2 cps   
β-channelapprox. 40 cps   
     
     
Typical efficiencies (ISO 7503-1)    
     
 LB 147 / LB 148Alpha Sense Version  
14C0,210,21  
36Cl0,540,71  
90Sr / 90Y0,430,64  
241Am0,190,4  
Typical detection limits (ISO 7503-1) with 10 s measuring time in Bq/cm2   
     
 LB 147 / LB 148Alpha Sense Version  
14C0.9540.961  
36Cl0.1910.137  
90Sr / 90Y0.2540.153  
241Am0.0550.026  