Sentra Biosains Dinamika

Berthold - LB 124 Alpha, Beta, Gamma Monitor

Catalog No. 60026-10 Category

PORTABLE CONTAMINATION MONITOR FOR ALPHA- AND BETA/GAMMA MEASUREMENT

The contamination monitor LB 124 SCINT Series, with simultaneous and separate measurement of alpha and beta-gamma radiation, are based on innovative scintillation technology. The compact and precise contamination measuring instruments are versatile and can be used flexibly in practical radiation protection.

The LB 124 SCINT Series have an attractive and ergonomic design and due to their low weights they are easy to handle. The measurement results can be easily read on a large high-resolution display with backlighting, even under unfavourable conditions.

General

  • Different detector sizes and types available
  • Light, handy and extremely rugged device
  • High sensitivity and uniform response
  • Calibration factors for over 50 nuclides
  • Calibration optionally according to ISO 7503-1 or with reference to an activity of 100 cm2

SCINT Technology

  • Unique simultaneous and separate measurement of alpha and beta-gamma radiation
  • The individual types of radiation can be distinguished by means of special evaluation and correlation circuits, and thus separated and measured at the same time
  • High sensitivity even at low energies
  • No counting gas required

Probes & Connectivity

  • Additional Dose Rate Detector (LB 124 SCINT-D)
  • Xenon filled proportional counter technology (LB 124 B)
  • RS232 interface
  • Headphone connection

Brand

Berthold

LB 124 SCINT
LB 124 SCINT-300
LB 124 SCINT-D
LB 124-B
ZnS detectorX
Xenon detectorXXX
Additional dose rate detectorXXX
Weight (in g)1300175014001620
Mechanical data
240 x 140 x 110
260 x 178 x 150
240 x 140 x 110
240 x 140 x 110
(L x W x D in mm)
Protection class (according to IEC 60529)IP 53IP 53IP 53IP 64
Sensitive detector area in cm²170345170150
Radiation detectorZnS(Ag) scintillatorZnS(Ag) scintillatorZnS(Ag) scintillator+ZnS(Ag) scintillator + Geiger-Müller tubesealed proportional counter tube with Xenon filling
Measurement modesSimultaneous alpha- and beta-gamma-measurement; ratemeter, scaler-timer-mode, clearance measurement, survey modeSimultaneous alpha- and beta-gamma-measurement; ratemeter, scaler-timer-mode, clearance measurement, survey modeα- and β-γ-measurement, simultaneous and separate, additional gamma dose rate H*(10). ratemeter, scaler-timer-mode, clearance measurement, survey moderatemeter, scaler-timer-mode, clearance measurement, survey mode, half-life measurement
Dimensions entrance window118 mm x 145 mm150 mm × 230 mm118 mm x 145 mm100 mm x 150 mm
Sensitive area170 cm²345 cm2170 cm²150 cm²
Material entrance window2 x 3 µm metallized plastic (0.4 mg/cm²)2 x 3 µm metallized plastic (0.4 mg/cm²)2 x 3 µm metallized plastic (0.4 mg/cm²)10 µm Titanium (approx. 5 mg/cm²)
Protective grid80 % transmission80 % transmission80 % transmission80 % transmission
Typical backgroundapprox. 10 cps
(α-channel)approx. 0.1 cpsapprox. 0.1 cps approx. 0.1 cps
(β-γ-channel)approx. 10 cpsapprox. 15 cpsapprox. 10 cps
Typical efficiencies (according
to ISO 7503-1)
241Amapprox. 44 %approx. 46 %approx. 44 %
14Capprox. 29 %approx. 30 %approx. 29 %
36Cl approx. 69 %approx. 69 %approx. 69 %
60Coapprox. 58 %approx. 58 %approx. 58 %
137Csapprox. 71 %approx. 74 %approx. 71 %
Gamma sensitivity at 1 µSv/h 137Cs
(α-channel)Not detectableNot detectableNot detectable
(β-γ-channel)< 100 cps< 100 cps< 100 cps
Spillover
α- in β-γ-channel< 20 %< 20 %< 20 %
β-γ- in α-channel< 2 x 10-5< 2 x 10-5< 2 x 10-5
Measuring range
(α-channel)0 to 5000 cps0 to 5000 cps0 to 5000 cps
(β-γ-channel)0 to 50000 cps0 to 50000 cps0 to 50000 cps
Gamma Radiation Detector
Dose rate range0.1 µSv/h to 20 mSv/h
Energy range50 keV to 1.3 MeV
Calibration factor0.652 µSv/h per cps 137Cs
Intr. backgroundapprox. 0.07 cps
Typical efficiencies (based on the activity of a source with 100 cm² area)
241Amapprox. 11 %
14Capprox. 2 %
36Cl approx. 24 %
60Coapprox. 26 %
137Csapprox. 20 %

BROCHURE