Investigating metabolic disorders with over 42 significant amino acids and their derivatives implicated in the analysis of Inborn Errors of Metabolism including Phenylketonuria (PKU).
This compact instrument allows users to accurately detect and quantify amino acids
from physiological samples in a fully automated system with an 84-position air, cooled autosampler, operate 24/7 ideal for busy clinical and research labs to analyze plasma, serum, urine, and dried blood samples.
Reproducibility Area | Better than 1.5% RSD at 10 nanomoles. Retention time: better than 0.5% RSD. |
Limit of Detection | 9 – 15 pmoles (primary amino acids detection at 570nm) |
Time of analysis | Accelerated analysis: 115 min injection to injection |
(based on the separation of 45 physiological amino acids) | High Performance analysis: 170 min injection to injection |
Analytical Column | High pressure PEEK column packed with Ultropac 8 cation exchange resin. Peltier heating/cooling system. |
Eluent System | 6 buffer system (5+1 regeneration solution) stored on the instrument at room temperature in graduated 1L glass |
bottles under nitrogen pressure. Ninhydrin reagent: Stored on the instrument at room temperature under nitrogen | |
pressure in a 2L plastic coated glass bottle. | |
Temperature | Column temperature variable between 20°C and 99°C. Reaction Coil temperature adjustable between 40°C and |
145°C (optimal temperatures: 135°C for UltraNin, 138°C for EZ Nin). | |
Photometric | Detection Single flow cell with optical beam splitter. Dual channel detection at 440nm and 570nm. |
Sample Injection | 3 injection modes (full loop, partial loop and micro), 84 position autosampler. Sample volumes from 1μL to |
5000μL. 200μL loop supplied as standard. | |
Software | BioSys control software |
Biochrom Alias Manager autosampler control software | |
OpenLAB CDS EZChrom Edition | |
Dimensions and Weights | Bench top fluidics cabinet: 48 x 59 x 57 cm, 19 x 23 x 22 inches (w x d x h) |
– Weight: 50 kg, 110 lbs | |
– Autosampler: 30 x 57.5 x 36 cm, 12 x 23 x 14 inches (w x d x h) | |
– Weight: 21 kg, 46 lbs | |
Operating Conditions | Operating temperature: 15 °C to 25 °C |
Maximum humidity: 80% at 25 °C | |
Required Services | Oxygen free nitrogen gas (99.99%) or Argon regulated to 73.5 psi (5bar). |
Drainage facility. | |
240V/100V, 50Hz/60Hz, 300 VA mains supply. | |
Safety System | Automatic shut-down and reaction coil flushing in the event of: |
• photometer lamp failure | |
• incorrect ninhydrin / buffer / coil / nitrogen pressures | |
• incorrect coil and column temperatures | |
• power failure |