Sentra Biosains Dinamika

Bio 30+ Series

Catalog No. 41562061 Category

Amino Acid Analyzer System, to monitoring and analyzing proteins, peptides and other ingredients in foods, beverages and livestock feed, to optimize nutritional value of your products, making them healthier to consume. The ability to accurately qualify and quantify amino acids at a molecular level is essential using gold standard techniques, the only one system established in the world. This system is exceptionally crucial to use in the lab where highest quality is to be achieved.  All chemicals and standards are available at highest quality control to support the AAA system worldwide.

Brand

Biochrom

Reproducibility AreaBetter than 1.5% RSD at 10 nanomoles. Retention time: Better than 0.5% RSD
Detection Limit9 – 15 pmoles (primary amino acids detection at 570nm)
Analysis TimeBiochrom 30+: Lithium System 115 minutes injection to injection
Biochrom 31+: Sodium citrate buffer system for hydrolysed proteins, 60 minutes injection to injection
Biochrom 32+: Sodium Accelerated Buffer System, 60 minutes injection to injection/Sodium Oxidised System,
90 minutes injection to injection.
Analytical ColumnHigh pressure PEEK column packed with Ultropac 8 cation exchange resin.
Peltier heating/cooling system.
Eluent SystemUp to 6 buffers (5+1 regeneration solution) stored on the instrument at room temperature in graduated
1L glass bottles under nitrogen pressure.
Ninhydrin reagent: Stored on the instrument at room temperature under nitrogen pressure in a 2L plastic
coated glass bottle
TemperatureColumn temperature variable between 20°C and 99°C. Reaction coil temperature adjustable between 40°C
and 145°C (138°C is optimum when using EZ Nin reagent).
Photometric DetectionSingle flow cell with optical beam splitter.
Dual channel detection at 440 nm and 570 nm
Sample Injection3 injection modes (full loop, partial loop and micro), 84 position autosampler (cooling optional)
Sample volumes from 1 μL to 5000 μL. 200 μL loop supplied as standard.
Software for system and autosampler control software
SoftwareData Handling software package—Openlab EZ Chrom Edition from Agilent
Dimensions and WeightsBench top fluidics cabinet: 48 x 59 x 57 cm, 19 x 23 x 22 inches (w x d x h)
– Weight: 50 kg, 110 lbs
– Autosampler: 30 x 57.5 x 36 cm, 12 x 23 x 14 inches (w x d x h)
– Weight: 21 kg, 46 lbs
Operating ConditionsOperating temperature: 15 °C to 25 °C
Maximum humidity: 80% at 25 °C
Required ServicesOxygen free nitrogen gas (99.99%) or Argon regulated to 73.5 psi (5bar).
Drainage facility.
240V/100V, 50Hz/60Hz, 300 VA mains supply.
Safety SystemAutomatic shut-down and reaction coil flushing in the event of:
• photometer lamp failure
• incorrect ninhydrin / buffer / coil / nitrogen pressures
• incorrect coil and column temperatures power failure