Sentra Biosains Dinamika

Bio 30+ Series

Catalog No. 41-5620-64 Category

Amino Acid Analyzer System

The gold standard dedicated Amino Acid Analyser for in vitro diagnostic use and research

A proven instrument for the precise quantification and qualification in free amino acids of physiological fluids,quantifying amino acids in inborn errors of metabolism, including Phenylketonuria, the technique complies with the new In Vitro Diagnostic Regulatory (IVDR) requirements. The Bio 30+ System is a gold standard dedicated amino acid Analyser used by reference laboratories, hospitals, pharmaceutical and industrial labs worldwide.

Brand

Biochrom

Reproducibility AreaBetter than 1.5% RSD at 10 nanomoles. Retention time: better than 0.5% RSD.
Limit of Detection9 – 15 pmoles (primary amino acids detection at 570nm)
Time of analysisAccelerated analysis: 115 min injection to injection
(based on the separation of 45 physiological amino acids)High Performance analysis: 170 min injection to injection
Analytical ColumnHigh pressure PEEK column packed with Ultropac 8 cation exchange resin. Peltier heating/cooling system.
Eluent System
6 buffer system (5+1 regeneration solution) stored on the instrument at room temperature in graduated 1L glass
bottles under nitrogen pressure. Ninhydrin reagent: Stored on the instrument at room temperature under nitrogen
pressure in a 2L plastic coated glass bottle.
TemperatureColumn temperature variable between 20°C and 99°C. Reaction Coil temperature adjustable between 40°C and
145°C (optimal temperatures: 135°C for UltraNin, 138°C for EZ Nin).
PhotometricDetection Single flow cell with optical beam splitter. Dual channel detection at 440nm and 570nm.
5000μL. 200μL loop supplied as standard.
SoftwareBioSys control software
Biochrom Alias Manager autosampler control software
OpenLAB CDS EZChrom Edition
Dimensions and WeightsBench top fluidics cabinet: 48 x 59 x 57 cm, 19 x 23 x 22 inches (w x d x h)
– Weight: 50 kg, 110 lbs
– Autosampler: 30 x 57.5 x 36 cm, 12 x 23 x 14 inches (w x d x h)
– Weight: 21 kg, 46 lbs
Operating ConditionsOperating temperature: 15 °C to 25 °C
Maximum humidity: 80% at 25 °C
Required ServicesOxygen free nitrogen gas (99.99%) or Argon regulated to 73.5 psi (5bar).
Drainage facility.
240V/100V, 50Hz/60Hz, 300 VA mains supply.
Safety SystemAutomatic shut-down and reaction coil flushing in the event of:
• photometer lamp failure
• incorrect ninhydrin / buffer / coil / nitrogen pressures
• incorrect coil and column temperatures
• power failure

APPLICATION

BROCHURE